User's Manual Part 2

6-29
1. Measurement Unit
2. Current Vital Sigh
3. Alarm Limits
4. OVR/UND for Temp extended range <30.0 °C or >44.0 °C(<86.0°F or >111.2°F ),
accuracy ± 0.4 °C
6.5.3. Temperature Patient Application
6.5.3.1. Surface Temperature Application
Perform the following procedure to apply the fiber-optic temperature sensor to a patient:
1. Carefully uncoil the sensor, avoid knotting or kinking the fiber optic cable
2. Inspect sensor for damage, including tears or deformations
3. Thoroughly clean and dry the axilla or groin application site
4. Position the sensor tip at the application site
5. Secure sensor with some medical tape
6. Cover application site to block air drafts
! WARNING
Avoid the use of any metalized foil temperature probe covers, or hydrogel backed probe
covers, as excessive RF heating could occur resulting in inaccurate temperature
measurements and/or burns.
NOTE
There is a temperature difference between surface temperature and patient body (core)
temperature.
When monitoring temperature during MRI procedures, the radio frequency (RF) energy
may normally increase the patient’s body temperature.
6.5.3.2. Setup Checklist
Temperature cable is properly inserted into the 3880 system
Temperature sensor is applied and positioned correctly
Temperature sensor is not damaged in any way
6.5.4. Changing the Temperature Settings
6.5.4.1. Units
The temperature format can be in Celsius or Fahrenheit. To adjust the units follow these steps:
1. Touch the Temperature vital sign box
2. Select Unit
3. Make your selection
4. Touch Back button to close menu