Problem
Possible Cause
Solution
EtCO2 Vital Signs are not
performing as expected
All other issues
1) Ensure the EtCO2
parameter is
enabled.
2) Inspect all EtCO2
components for
damage and leaks
and replace as
necessary.
3) Review section 6.3
of the operators
manual.
4) Contact your local
representative to
schedule
applications training.
4
NIBP values seem
unstable
Compromised hose or
cuff
Check that tubing is not
stretched, bent,
compressed or loose.
Ensure the patient is not
shivering or moving.
Ensure the correct sized
cuff is being used and
applied correctly.
Replace Hose or cuff if a
leak is suspected.
Monitor has an internal
leak
Refer to qualified technical
service representative.
Patients Arm Position
Ensure that the patients
arm is at heart level.
Cuff comes off patient
limb
Inappropriate sized cuff
Determine correct cuff
size and apply to patient
according to section 6.
Cuff applied inside out
Reapply cuff according to
section 6.
NIBP not functioning
NIBP hose or cuff is
disconnected
Ensure NIBP hose is
securely connected.
Monitor has an internal
failure
Refer to qualified technical
service representative.
NIBP Vital Signs are not
performing as expected
All other issues
1) Ensure the NIBP
parameter is
enabled.
2) Inspect all NIBP
components for
damage and leaks
and replace as
necessary.
3) Review section 6.4
of the operators
manual.
4) Contact your local
representative to
schedule
applications training.