User's Manual Part 2
8-17
Problem
Possible Cause
Solution
All other issues
1) Ensure the ePOD is
on the same
wireless network
channel as the 3880.
2) Ensure the ECG
parameter is
enabled.
3) Inspect all ECG
components for
damage and replace
as necessary.
4) Review section 6.1
of the operators
manual
5) Contact your local
representative to
schedule
applications training.
2
SpO2 not functioning
SpO2 sensor is not
attached to the patient.
Ensure SpO2 sensor is
securely attached.
Poor SpO2 sensor
placement
Reposition/replace SpO2
sensor according to
section 6.
Weak radio link
Verify radio channel
setting and signal level.
Reposition antenna for
improved reception.
Faulty SpO2 oPOD
Replace SpO2 oPOD, or
refer to qualified technical
service representative.
SPO2 readings are
unstable or poor
Sensor positioning
Check the P.I. value and
try repositioning the
sensor correctly on the
patient according to
section 6.
Ensure the SpO2 cable is
correctly inserted into the
SpO2 applicator / grip.
Averaging time
Try adjusting the SpO2
averaging time.
Patients specific anatomy
Skin pigment and certain
anatomy ailments can
cause inconsistent SpO2
readings.
Check that there is no nail
polish on the patients digit
impeding the readings.
Cool temperature is
affecting patients
perfusion
Check the P.I. value and
ensure the patient is not
shivering and that their
digits are warm to the
touch.