User's Manual Part 2
8-6
8.5. User Maintenance
8.5.1. Overview
Before every use, a thorough inspection should be performed.
Follow these guidelines when inspecting the 3880 monitor:
• Make sure that the environment and power supply meet the requirements.
• Inspect the 3880 system and its accessories for mechanical damage.
• Inspect all power cords for damage, and make sure that their insulation is in good
condition.
• Make sure that only specified accessories are applied.
• Inspect if the alarm system functions correctly.
• Make sure that the recorder functions correctly and the recorder paper meets the
requirements.
• Make sure that the patient monitor is in good working condition, and batteries have
sufficient charge.
In case of any damage or abnormity, do not use the patient monitor. Contact the hospital’s
biomedical engineers or your authorized
IRadimed service personnel immediately.
8.5.2. Updating Software
As revisions of the software become available, the 3880 can be updated, see service manual
for detail procedure.
8.6. Troubleshooting Problems With No Onscreen Message
The following section is for troubleshooting the 3880 system when there is no corresponding
message displayed on the screen. If a message is displayed on the screen please review Exhibit B of
the operators manual. Please contact IRadimed Technical support at (407) 677-8022 if you need
assistance troubleshooting or are unable to resolve any issues.
8.6.1. Troubleshooting Power Related Problems
Problem
Possible Cause
Solution
1
3880 will not turn on.
Power dial is not in the
correct position
1) Turn 3880 ON by
rotating the power
dial to the ON
position.
No AC power/battery
depleted.
1) Plug AC Adapter
power cord into a
working AC outlet
immediately.
2) Plug AC Adapter
into the circular
power receptacle on
the rear of the
device.
3) Optional Swap low
battery with a fully
charged battery.