8.4. Care and Cleaning
8.4.1. Introduction
For safe and reliable function and operation of the monitor, regular care has to be carried out in
accordance with manufactures guidelines. Use only manufactures approved substances and
methods to clean and disinfect your equipment. The warranty does not cover damage caused
by unapproved substances or methods.
IRadimed makes no claims regarding the efficacy of the listed chemicals or methods as a
means for controlling infection.
NOTE
If the monitoring system does not function as it should and troubleshooting cannot solve
the problem, contact your service representative.
Detailed cleaning instructions provided with any accessory should be followed.
8.4.2. General Guidelines
Keep your equipment and accessories free of dust, dirt and contaminants. To avoid damage to
the 3880 monitor follow these rules:
Always dilute liquid cleaning agents according the manufacturer’s instructions or use
lowest possible concentration.
Do not immerse part of the 3880 monitor into liquid.
Do not pour or use pressurized spray liquid onto the 3880 monitor or accessories.
Do not allow liquid to enter the case.
Never use abrasive materials, solvent or corrosive cleaners.
Clean the 3880 monitor in a well-ventilated area before and after each patient use.
Allow components to dry completely prior to use.
A soft bristled brush may be used to clean narrow areas.
Do not use hard or pointed objects to clean any part of the monitor.
! WARNING
Be sure to shut down the system and disconnect all power cables from the outlets
before cleaning the 3880 monitor.
Do not use unspecified cleaners, materials or methods as they may damage the device,
labels or cause failures.
Do not use conductive solutions or materials to clean the system.
Do not reuse sensors intended for single patient use.
! CAUTION
If liquid is spilled on the 3880 monitor or accessories and may have entered the system
immediately contact your service personnel.
Do not use solvent based cleaning agents as damage to the plastic parts could occur.
NOTE
Before cleaning any equipment, consult your hospital’s regulations for cleaning
Refer to any superseding instructions accompanying any accessories or options.