User's Manual Part 2
8-3
• Occlude the sampling line and check that the message “Occlusion” appears within 30
seconds
NIBP
• Check that the pressure values are displayed for all cuff sizes
Temperature
• Check that the temperature value is displayed when the cable is connected
Recorder
• Check that the strip chart recorder is functional
Alarms
• Turn the monitoring system on and check that the Red, Amber and blue Tri-Color Tri-
Color Alarm Dome Light are lit momentarily
• Check that the speaker gives and audible tones at the desired level
• Use a conventional patient simulator and / or test gasses to verify alarm functionality.
When a problem with the alarm system is suspected, always refer the monitor to a
qualified service personnel.
Battery
• The battery pack should be inspected anytime the battery pack is removed. Look for
signs of physical damage, shock and swelling. Check that the battery pack is holding
sufficient operating capacity.
• Check that the battery is communicating with the monitor by observing the battery icon
on the display.
! CAUTION
• Any failure of the inspection of the battery pack will require discontinuance of use and
replacement will be needed.
• If any cells swollen greater than 8 mm thick or cause the plastic case to bulge
discontinue use and contact your support personnel for proper disposal.
8.2.3. Every Twelve Months
Preventative Maintenance Check
• Calibration check of temperature, NIBP and CO
2
/ Multi-Gas Agents Unit
NOTE
• The annual check according to detailed instructions of the Service Manual requires
trained service personnel and appropriate testing tools and equipment.
• Verify all calibration / test gas mixture containers are empty before disposal.
• Use only MRI compliant calibration / test gas mixture containers suitable for use in the
MRI environment for verification of gas readings inside of zone IV.
8.3. Service Setup Menu
8.3.1. Software Version
To check the current revision of software in the 3880 system, PODs and Tablet follow these
steps:
1. Press the SETTINGS Button
2. Select “Service Mode”
3. Enter password
4. Select “Device Information”
5. Check software version
6. Touch Back button to close the menu