User's Manual Part 1
2-41
3880 multi-parameter monitor
The 3880 system has a user replaceable battery located on the rear of the device.
1. Insert the user replaceable battery pack into the battery bay on the rear of the 3880
monitor as shown. The battery pack will lock into place if inserted properly.
2. Connect the system to AC power. Reference the Charging Instructions section 2.2.7.2 in
this manual for additional details.
3. Verify that a green or yellow LED is present on front panel indicating mains power
(Green) and charging (Yellow).
4. Allow to charge for a minimum of 5 hours prior to initial use.
3881 ECG and 3882 SpO
2
PODs
The 3880 PODs have an integrated battery that is not user serviceable. Additional PODs
maybe added to this system to allow for continuous operation on battery. The 3880 Non-
Magnetic patient monitor has the ability to charge 2 PODs simultaneously.
1. Disconnect the Patient and insert the PODs with the labels facing outward into the
charging bays located on either side of the 3880 monitor.
2. Connect the system to AC power. Reference the Charging Instructions section 2.2.7.2.1
in this manual for additional details.
3. Verify associated charge indicator LEDs on rear panel.
4. Allow to charge for a minimum of 4 hours prior to initial use.