User's Manual Part 1
2-38
2.2. Initial Setup
2.2.1. Safety Precautions
! WARNING
• The monitor and its components should not be used adjacent to or stacked with non
IRadimed equipment. If adjacent or stacked use is necessary, the monitor and its
components should be observed to verify normal operation in the configuration prior to
use. Never install the monitor directly above the patient.
• To avoid risk of electric shock, this equipment must only be connected to properly
grounded AC outlets with the power cord supplied by IRadimed.
! CAUTION
• If the monitor is going to be in storage and not used for more than 30 days, the battery
pack should be removed from the devices if possible and charged to 50%.
• If the PODs are going into storage for more than 30 days they should be stored 50%
charged and recharged at a minimum of 6 month intervals.
2.2.2. Unpacking the System
1. Confirm that the packing box is undamaged. If the box is damaged, contact the shipper.
2. Open the top of the box and carefully unpack all components.
3. Confirm that all components are undamaged. If any of the components are damaged,
contact the shipper.
4. Confirm that all components are included. If any of the components are missing, contact
your
IRadimed representative.
a. Standard Equipment:
• 3880 Monitor
• ECG and
SpO
2
PODs
• 1120 Power Supply and Cables(P/N 1121, 1122)
• Operators & Service Manual
• ECG,
SpO
2
and NIBP accessories
b. Optional Equipment:
• 3885-T Remote Tablet and 3885-B Base Station
• Hi-Gain Antenna
• CO
2
, Gas, Fiber Optic Temperature, and Recorder accessories
• Gating Cable
• Mounting Hardware and Stand
! CAUTION
• Never unpack or assemble items inside the MRI room suite (Zone IV).