User's Manual Part 3

10-24
E.3.2. Gas Analyzer
Sensor head
2 to 9 channel NDIR type gas analyzer measuring at 4 to 10 µm. Data
acquisition rate 10 kHz(sample rate 20 Hz/channel)
Compensations
ISA OR+:
Automatic compensation for pressure and temperature.
Manual compensation for broadening effects on CO
2
Calibration
No span calibration is required for the IR bench. An automatic Zeroing is
performed typically 1 to 3 times per day.
Warm-up time
ISA OR+:
<20 seconds (Concentrations reported, automatic agent
identification enabled and full accuracy)
Rise time
1
At 50 sml/min sample flow
ISA OR+
CO
2
300 ms
N
2
O, O
2
, ENF, ISO, SEV, DES < 400 ms
HAL < 400 ms
Primary agent threshold
(ISA OR+)
0.15 vol% When an agent is identified, concentrations will be reported even
below 0.15 vol% as long as apnea is not detected
Secondary agent threshold
(ISA OR+)
0.2 vol% + 10% of total agent concentration
Agent identification time
(ISA OR+)
< 20 seconds (typically < 10 seconds)
Total system response time
ISA OR+ < 4 seconds
(with 2 m Nomoline Airway Adapter Set sampling line)
1
Measured according to EN ISO 80601-2-55
E.3.3.
Accuracy - standard conditions
The following accuracy specifications are valid for dry single gases at 22 ± 5 'C and 1013 ± 40 hPa:
Gas
Range
1
Accuracy
CO
2
0 to 15 vol%
±(0.2 vol%+ 2% of reading)
N
2
O
0 to 100 vol%
±(2 vol% + 2% of reading)
HAL, ENF, ISO
0 to 8 vol%
±(0.15 vol% + 5% of reading)
SEV
0 to 10 vol%
±(0.15 vol% + 5% of reading)
DES
0 to 22 vol%
±(0.15 vol% + 5% of reading)
O
2
0 to 100 vol%
±(1 vol%+ 2% of reading)