User's Manual Part 3

8-16
Problem
Possible Cause
Solution
Incompatible lead view
for the MRI scan and/or
electrode placement
Change the ECG lead view
to select the best
performing view for the
sequence.
Poor skin preparation
Re-prep the patient
according the application
procedure found in
section 6.
Poor electrode
placement
Reposition/replace
electrodes according to
the application procedure
found in section 6.
Poor electrode quality
Replace the electrode with
an IRadimed approved
electrode.
Weak radio link
Verify radio channel
setting and signal level.
Reposition antenna for
improved reception.
QRS amplitude is less
than 10 mm/mv
Change the lead view and
/ or reprep and reposition
the electrodes.
Faulty ECG ePOD
Replace ECG module, or
refer to qualified technical
service representative.
Incorrect ECG Mode
selected
Select the appropriate
ECG filter mode for the
application. Refer to
section 6 of the operators
manual.
Patient motion
Ensure that the patient is
not shivering or moving.
ECG lead not properly
inserted into the ePOD
Ensure each ECG leadwire
is inserted completely into
the ePOD.
Damaged ECG lead
Replace ECG lead with an
IRadimed approved ECG
lead.