User's Manual Part 3

8-14
Problem
Possible Cause
Solution
MRI Room Attenuation.
1) Ensure the high gain
antenna is attached
to the correct
receptacle on the
3885-B
2) Position the high
gain antenna so it
has a direct line of
sight to the 3880.
3) Try a different
wireless channel
4) Contact IRadimed
Technical Service.
5
ePOD or oPOD will not
communicate with a
3880
3880 is not turned ON.
1) Ensure the 3880 is
turned on and is
operational.
3880 is not within
wireless range.
1) Verify PODS are
within 9 ft (3 m) of
3880 and that no
other devices, walls
or doors are blocking
the wireless signal.
2) Reposition 3880 to
have a line of sight
to the PODS as
needed
Incorrect Channel.
1) Ensure that the
ePOD, oPOD and
3880 are operating
on the same wireless
network channel.
2) Record wireless
network from the
3880.
3) Select matching
wireless channel on
3881 ePOD
4) Select matching
wireless channel on
3882 oPOD.
Multiple system
components on the same
wireless network
channel.
1) Ensure that each
3881 and 3882 in
your facility are on
separate, unique
channels that match
a corresponding
3880 (or not used
simultaneously).
6
Printer will not print
Wireless communication
to the 3885-B
1) Ensure wireless
connection is
established to a
3885-B equipped
with a recorder