User's Manual Part 3
8-9
Problem
Possible Cause
Solution
AC power source has
incorrect voltage.
1) Switch AC power
cord to a power
receptacle with
sufficient voltage.
Refer to the service
manual for voltage
requirements.
5
ePOD or oPOD will not
turn on
Power ON key not
pressed.
1) Toggle the power
switch to the ON
position.
Listen and feel for
the tactile feedback
to confirm the button
has been properly
pressed.
Insufficient battery life
1) Plug 3880 system
into AC power.
2) Ensure the POD is
correctly docked and
seated into a
charging bay on the
3880 monitor.
3) Allow to charge a
minimum of 4 hours.
4) Contact IRadimed
Technical Support
6
3880 will not operate
on battery.
Battery not properly
charged.
1) Plug AC Adapter
power cord into an
AC outlet
immediately.
2) Plug AC Adapter
into the circular
power receptacle on
the rear of the
device.
3) Wait until battery is
fully charged.
4) Press the button on
the exterior of the
battery pack. If no
LED’s illuminate
replace the battery
pack before use.