User's Manual Part 2

2-58
2.2.12.4.2 Adjusting the 3881 ECG ePOD Wireless Channel
1. Identify the channel setting of the 3880 monitor that the ePOD should communicate to
2. Power on the wireless ePOD
3. Press the yellow CH Select” button to advance to the next wireless communicating
channel in sequence, choosing the channel which matches the 3880monitor
a. A white channel LED will light for channels 1(5), 2(6), 3(7) or 4(8)
b. A blue Channel Select” LED will illuminate indicating a shift to channels 5, 6, 7
and 8
4. Confirm that the ePOD is communicating with the 3880 system by observing the battery
icon on the screen. And white connected LED on ePOD.
2.2.12.4.3 Adjusting the 3882 SpO
2
oPOD Wireless Channel
1. Identify the channel of the 3880 monitor that the oPOD should communicate to
2. Power on the wireless oPOD
3. Press the yellow “CH Select” button to advance to the next wireless communicating
channel in sequence, choosing the channel which matches the 3880monitor
a. A white channel LED will light for channels 1(5), 2(6), 3(7) or 4(8)
b. A blue Channel Select” LED will illuminate for channels 5, 6, 7 and 8
4. Confirm that the oPOD is communicated with the 3880 system by observing the battery
icon on the screen. And white connected LED on oPOD.
2.2.12.4.4 Adjusting the 3885-T Remote Tablet Wireless Channel
1. Identify the channel of the 3880 monitor that the 3885-T Remote Tablet should
communicate to
2. Access the Wireless Setup Screen
a. Touch the wireless channel icon in the top left corner
3. Select the intended wireless channel (identified by word “Channel” followed by a
number)
4. Confirm that the 3885-T Remote Tablet is communicated with the 3880 system by
observing the battery icon on the screen and the corresponding LED flashes.