User's Manual Part 2
4-8
4.4. Controlling Alarms
4.4.1. Accessing Alarm Menu
Quick Alarm setup allows a targeted adjustment of 1 vital sign measurement. The “quick alarm”
is accessed by a user by pressing the corresponding vital sign box.
1. Lower Alarm Limit Fine Adjustment
2. Lower Alarm Limit Quick Slide Adjustment
3. Upper Alarm Limit Quick Slide Adjustment
4. Upper Alarm Limit Fine Adjustment
NOTE
• Green: range of numerical values that are inside the alarm limits and will not trigger an
alarm.
• Red: range of numerical values that are outside the alarm limits and will trigger an
alarm.
• The system automatically prevents the crossover of High and Low Limit settings.
4.4.2 Gas (Anesthetic Agent) Alarm Screen
1. Nitrous Oxide (N
2
O) Inspired Limit Selector, high limit cannot be set above 80%
2. Halothane (Hal) Expired Limit Selector
3. Enflurane (Enf) Expired Limit Selector
4. Isoflurane (Iso) Expired Limit Selector
5. Sevoflurane (Sev) Expired Limit Selector
6. Desflurane (Des) Expired Limit Selector
7. Oxygen (O
2
) Inspired Limit Selector, cannot be set below 18% without special key
8. Halothane (Hal) Inspired Limit Selector
8
7
4
5
6
3
2
1
12
11
10
9
4
3
2
1