User's Manual Part 2

3-1
3. Advanced Case Management Strategies
The 3880 system includes several features to help facilitate the efficient management of
patients undergoing a MRI procedure.
3.1. Case Management
Many of the case management strategies are intended for use with a fully equipped 3880
system including the 3885-B Base Station and 3885-T Remote Tablet. Various strategy
suggestions provided in this document are examples showing the capabilities of the system.
Always consult local policy prior to initiating any strategy into clinical practice.
3.1.1. Preparing for a patient
Prior to starting the case certain monitor features can be prepared.
Proper use of the ALARM OFF /STANDBY feature is an efficient way to suspend
alarms and NIBP patients in between patients
Configure the monitor with the correct settings and patient worn accessories
appropriate for the case
Edit the patient identifiers specific to the case
3.1.1.1. Clearing previous data
If trend data is still stored inside the monitor from a previous patient it can be erased by
following these steps.
1. Press the Trends hard key button
2. Select the “CLEAR” Trends soft key button
3. Select Yes to confirm
4. Touch the Back button to exit
3.2. Multiple System Wireless Strategies
The 3880 systems extended range wireless can help improve the efficiency for environments
that want to improve MRI throughput. MRI rooms that utilize multiple 3880 monitors with a
single Base station can gain efficiencies through monitor rotation. When owning multiple
systems ensure that each 3880 has a unique wireless channel. Additionally each installed
3885-B Base Station cannot share a wireless channel with another Base Station within range,
at the facility. Set the 3885-T Tablet and 3885-B Base to the same wireless channels to match
the 3880 monitor that the patient arrives on.
MRI
ROOM
Wireless
Channel
1
Wireless
Channel
2
Alternate base and Tablet
between Wireless Channel 1
and 2 depending on which
3880 is in the MRI room.
Verify Base and Tablet are on
the SAME channel