User's Manual Part 1
2-44
3885-B Base Station, Optional
1. Locate the power cord (P/N 1128 in US) and ensure the Base Station power switch is in
the OFF position. (1)
2. Locate a suitable location in the control room that enables access and wireless
communication. See 2.2.8 for antenna connection.
3. Attach the power cord to the 3885-B Base Station (2)
4. Place the power cord into the designated AC power outlet.
5. Turn the 3885-B Base Station power switch to the ON position (1)
6. Check for green LED light on the front of the 3885-B Base Station.
! WARNING
• The 1120 power supply contains magnetic properties. Keep outside the 1,000 Gauss
line, or 10 feet (3 meters) from the MRI system.
• Never Velcro or attach the power supply directly to the monitor, monitor mounting
solution, or IV pole.
• The 3885-B Base Station contains ferrous materials that are attracted to magnetic fields.
DO NOT install or place the 3885-B Base Station or its components in the MRI room
(Zone IV).
2.2.8. Antenna Strategies
The 3880 system uses 2.4 GHz bidirectional communication to communicate between the 3880
monitor and the 3885-T Remote Tablet. There are different antenna options to support various
environments and workflow scenarios. Please contact your IRadimed representative for
information regarding one of the optional antenna options.
2.2.8.1. Omni Directional Antenna
The Omni Directional Antenna is the standard antenna that is included with the 3880 Monitor
and 3885-B Base Station. This bidirectional broadcast antenna is the easiest to use and
provides patient workflow flexibility in typical MRI environments.
! CAUTION
• Locate the 3885-B Base antennas, the 3880 and 3885-T and the PODs more than 1
MRI Room