User's Manual Part 1
1-7
• To protect against injury, follow the directions below:
o Avoid placing the device on surfaces with visible liquid spills.
o Do not soak or immerse the device in liquids.
o Do not attempt to sterilize the device.
o Use cleaning solutions only as instructed in this operator's manual.
Do not attempt to clean the device while monitoring a patient.
Do not submerge the pulse oximeter oPOD or ECG ePOD in any cleaning solution or attempt to
sterilize by autoclave, irradiation, steam, gas, ethylene oxide or any other method. This will
seriously damage the pulse oximeter oPOD or ePOD.
Contraindication: Use of ECG monitoring is contraindicated on patients with conductive or
active implant devices, including pacemakers or electrical stimulators.
1.3.2. MRI Monitor
Perform operational checkout before each use. If monitor fails to function properly, refer to
qualified service personnel.
For safe and accurate operation and MRI compatibility, use only recommended
IRadimed
Corporation patient electrodes, cabling, lead wires, cuffs, hoses, sensors, tubing, etc. A listing
of these can be found in the Accessory Listing within this manual, or by contacting
IRadimed
Corporation directly.
For continued operation, always connect the monitor to AC Main Power through the 1120 AC
Power Supply when a Low Battery indication occurs. Failure to do this can lead to interruption
of monitoring.
The system may not conform to all performance specifications if stored or used outside the
environmental specifications identified in Exhibit A in the rear of this manual.
Do not apply any unnecessary pressure to the screen area of the monitor. Severe pressure
applied to this portion of the monitor could result in damage or failure of this screen.
All equipment not complying with IEC 60601-1 should be placed outside the patient
environment. Only connect IEC 60601-1 compliant equipment to this monitor. To avoid
potentially hazardous leakage currents, always check the summation of leakage currents when
several items of equipment are interconnected.
For proper equipment maintenance, perform the service procedures at the recommended
intervals as described in the monitor’s service manual.
Single use devices should never be reused. Risk of infection or inaccurate readings may result.
Do not use this system or accessories inside a hyperbaric oxygen chamber.
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.
Monitor is not intended for long-term data storage.
Do not press on the buttons or touchscreen with sharp or hard objects which could damage the
keys or screen. Use only finger tips with the buttons and screen. Do not place the 3880 MRI