Users Manual Part 7
Clinical User’s Guide
IBA | D-3|
Volume 1 - Treatment Session
Managing adaPT
prescribe
Settings
|
The Tolerance table is called when you define the beam, in standalone mode.
Performing Functions With Tolerance Tables
From the TOLERANCE TABLE MANAGEMENT SCREEN you can perform any of the
following:
View the existing tolerance tables.
Create a new tolerance table: the tolerance table label must be unique.
Making changes to the tolerance tables may have important consequences on
clinical operations.
Important: The treatment center will issue and enforce a procedure requesting
the responsible medical physicist to print, verify, sign, and file the list of software
stored Tolerance tables at specific intervals.
Internal Tolerance Tables
When you click an internal tolerance table, the buttons at the bottom of the
T
OLERANCE TABLE MANAGEMENT SCREEN enable you to perform the following
functions:
Copy: you can copy all the values of the existing tolerance table, except the
tolerance table label. You have to enter a new and unique label before being
allowed to save the table.
Edit: you can edit an existing tolerance table only if the tolerance table has never
been linked to a setup or treatment beam.
Delete. you can delete an existing tolerance table only if the tolerance table has
never been linked to a setup or treatment beam.
WARNING Any changes made to tolerance tables may cause death, bodily injury,
or poor therapeutic results if not performed correctly.