Users Manual Part 7
IBA | F-14 |
Clinical User’s Guide
Volume 1 - Treatment Session
System Messages
|
adaPT-
deliver
Any Any Too many set range failures due to field
regulation. Please inform the Service
team. Sign in if you want to disable field
regulation and continue irradiation in
degraded mode.
The system has been unable to correctly reach the
expected magnetic fields for some magnets (for
which automatic field regulation is applicable) after
3 consecutive attempts. As a result, it is
recommended to the user to switch the system into
a "degraded mode" for which the field regulation is
not activated or to request for a verification of the
equipment by the service team.
adaPT-
deliver
Any Any The 'beam on' consistency check on
ICcyclo failed. Please inform the Service
team.\nIf you agree to override it, a
redundant beam check might be
ineffective.
When some beam is being extracted from the
accelerator, a discrepancy has been detected
between the beam presence signals provided by
the Ion Chambers located at the cyclotron exit (IC
cyclo: small gap and large gap). Such an discrepancy
indicates a potential failure of one of these Ion
Chambers and requires a verification by the service
team. Nevertheless it is possible to proceed with
the current treatment by means of an "override"
procedure (using a specific password). As long as
the issue is not fixed, a message shall inform the
user of this issue prior to each treatment.
Table F-1. System messages
Software Environment(s) Delivery Message Description