Users Manual Part 7

Clinical User’s Guide
IBA | F-13 |
Volume 1 - Treatment Session
System Messages
|
adaPT-
deliver
DevC Any The system encountered an issue
recording the irradiation in the
OIS.Please refer to the Dose Counter
Electronic Unit and manually record this
irradiation in the OIS.
This message informs the user that the system
could not record the irradiation in the OIS and it
instructs the user to manually record the irradiation
in the OIS using the data on the DCEU.
adaPT-
deliver
Standalone Any The system encountered an issue
recording this irradiation in the local
database.Please refer to the Dose
Counter Electronic Unit and manually
record this irradiation in the patient file.
This message informs the user that the system
could not record the irradiation in the database and
it instructs the user to manually record the
irradiation in the patient file using the data on the
DCEU.
adaPT-
deliver
Physics Mode Any You logged into physics mode. Clinical
treatments are not allowed in this mode.
This message informs the user that they have
logged into adaPTdeliver in Physics mode and that
patient treatment is not allowed in this mode.
adaPT-
deliver
Any Any Warning, the system (seems to)
encounter a problem, it is recommended
to stop the treatment and to contact an
IBA Technician!
This message indicates that, for the third time
within the last 15 seconds, a treatment interruption
(due to a failed periodical check) has occurred. In
such a case, it is strongly recommended to the user
to terminate the treatment.
adaPT-
deliver
Any Any Irradiation done. Please contact beam
operator to perform a new RF lookup.
This message indicates that the last treatment has
been performed in such conditions that require the
performance of a new "RF lookup" sequence prior to
starting the next treatment. The user must contact
an operator to complete the task needed.
Table F-1. System messages
Software Environment(s) Delivery Message Description