Users Manual Part 7

Clinical User’s Guide
IBA | F-11 |
Volume 1 - Treatment Session
System Messages
|
adaPT-
deliver
OIS Any Unable to send treatment record to OIS
and failed to save local disk.
This message informs the user that the system
could not record the irradiation in the OIS or save a
treatment record in the local disk. In this case, the
user should refer to the Dose Counter Electronic
Unit to manually record the irradiation in the OIS.
adaPT-
deliver
Any Any This beam has been partially irradiated;
are you sure you do not want to enter a
previously delivered MU?
This message informs the user that, despite the fact
that the current beam has been partially irradiated in
the past, they have not entered an MUd value to
resume from. The message prompts them to
confirm whether this corresponds to their needs for
the current irradiation.
adaPT-
deliver
Any Any Unable to save the current setup record
into database due to database access
exception. You are recommended to
check database connection and abort
this session.
This message informs the user that the current
setup could not be recorded in the database. The
message recommends that they do not continue
with the session and that they fix the issue (with
IBA support if necessary) before proceeding with
treatment.
adaPT-
deliver
DevC
Standalone
Any No setup beam has been delivered yet.
Do you confirm selection of the
treatment beam?
This message prompts the user to confirm the
selection of a treatment beam, bringing to their
attention that no setup beam has been delivered in
the current session.
Table F-1. System messages
Software Environment(s) Delivery Message Description