Users Manual Part 7

IBA | F-10 |
Clinical User’s Guide
Volume 1 - Treatment Session
System Messages
|
adaPT-
deliver
Any Any The irradiation has stopped abnormally.
The reason of the termination is
unknown.\n[Treatment Termination
Status={0}]
This message informs the user that the irradiation
was stopped before being complete, due to an
unknown reason. The message also provides the
Treatment termination Status of the irradiation.
adaPT-
deliver
Any Any PMS seems busy. Do you want to start
preparation anyway?
This message informs the user that the PMS is busy
and to confirms whether the system should start
the preparation anyway.
adaPT-
deliver
DevC Any Unable to store the plan data because:{0} This message appears when the plan cannot be
saved into the database.
adaPT-
deliver
DEVC Any The system encountered an issue
recording the session in the OIS. A
treatment record has been saved to local
disk:{0} . Please collect the treatment
record file saved on disk and manually
import it in the OIS.
This message informs the user that the system
could not record the session in the OIS and that it
saved a treatment record in the local disk. the
message instructs the user to manually import the
treatment record saved on disk in the OIS.
adaPT-
deliver
OIS Any The system encountered an issue
recording the irradiation in the OIS.
Please refer to the Dose Counter
Electronic Unit and manually record this
irradiation in the OIS.
This message informs the user that the system
does not have information about the dose to record
the irradiation in the OIS. The message instructs the
user to refer to the Dose Counter Electronic Unit to
manually record the irradiation in the OIS.
Table F-1. System messages
Software Environment(s) Delivery Message Description