Volume 1 - Treatment Session
System Messages
|
Table F-1. System messages
Software Environment(s) Delivery Message Description
Any DEVC Any Failure to store Dicom object to Dicom
server.SopClassUID={0}SopInstanceUID
={1}
This message informs the user that the system has
failed to store a DICOM object into the DICOM
server. Following this message, the user may need
to manually record the treatment.
Any DEVC Any This prescription has been partially
delivered but detailed delivery history
cannot be fully retrieved. As a result,
partial information will need to be
manually cross-checked.
This message informs the user that they need to
manually cross-check the information related to a
partial irradiation, as the system could not retrieve
its detailed delivery history.
Any DEVC Any A Treatment record was referenced but
could not be retrieved.UID={0}, user
must check the tx history to manually
enter partial information, Do you want to
continue?
This message informs the user that they need to
check the treatment history and manually enter
partial information if they want to continue with an
irradiation for which the system could not retrieve a
treatment record.
Any DEVC Any Do you really want to cancel current
appointment?
This message prompts the user to confirm they
wish to cancel the selected appointment and close
the session.
Any DEVC Any OIS provides incomplete data for the
computation, the dose limits check will
be ignored.
This message informs the user that, as the OIS has
provided data that is not sufficient for the system to
perform dose limit checks, these checks will not be
performed.