Users Manual Part 7
Clinical User’s Guide
IBA | E-1 |
Volume 1 - Treatment Session
Appendix E
• • • • • •
Emergency Stop Button Locations
Table D-1 lists all emergency stop buttons and the Safety Redundant Control Unit
(SRCU) to which they are connected.
Figure D-1 illustrates the location of all safety and emergency devices.
Note: The switch depicted on the hand-pendant represents the pressure sensitive
switches at the bottom of the hand pendant that enable or disable the movement of
the Patient Positioning Devices (refer to the Clinical User's Guide for the use of the
hand-pendant).
Table E-1. Emergency Stop Button Data
Location Quantity Operated by
Main Control Room 1 SRCU0
Cyclotron & Beam Transportation Vault
Subzone 1 5
Subzone 2 5
Gantry Treatment Room 1
Vault 4 SRCU1
Catwalk 2
X-ray Booth 1
Treatment Room 5
Treatment Control Room 1
PPS maintenance hand pendant; no electrical power shut off 1