Users Manual Part 6

IBA | 49-12 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part XI
- Patient Positioning and Alignment Aligning a Patient in the GTR
|
Confirming the Treatment Position
When you are satisfied that the Treatment position is correct, confirm it as follows:
1. Touch the green (Tick) icon at the top right of the wireless hand-pendant
screen.
Note: If the current position is outside the treatment volume, the Position
outside treatment volume. Position verifications are
mandatory message appears on the hand-pendant.
In case the current position is out of tolerance or outside the treatment volume,
this OK acknowledges that you accept the situation. If you do not want to accept
the situation, press Cancel.
The system compares the values of the required setup position against limits defined
in the tolerance tables. If the check is successful, the corrected position shall be
saved, the patient position status Treatment Saved check box is selected on the TR
monitor.