Users Manual Part 6

Clinical User’s Guide
IBA | 49-11|
Volume 1 - Treatment Session
|Part XI
- Patient Positioning and Alignment Aligning a Patient in the GTR
|
Implementing the Position Corrections
Use the wireless hand-pendant or remote positioning controls in Auto mode to
implement the corrections by moving the PPS to the Corrected target position, as
described in Chapter 9, “Using the Wireless Hand-Pendant”, and Chapter 11, “Using
the Remote Positioning Controls”, respectively.
When the PPS reaches the Corrected position, the hand-pendant or the remote
positioning controls beep one time.
If the remaining device(s) is (are) at the required position when the PPS reaches the
Corrected position the hand-pendant beeps two times.
WARNING Make sure that the DID flat panel arms are fully retracted before
attempting to move Patient Positioning Devices (PPDs) that can collide
with the flat panels.
WARNING After applying and implementing the corrections calculated by the
Patient Position Verification System (PPVS) software (e.g.,
adaPTinsight), it is recommended to take a new set of X-ray images to
verify proper positioning of the patient.