Users Manual Part 6

IBA | 48-8 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part XI
- Patient Positioning and Alignment Using the Universal Beam Triggering Interface
|
Figure 48-2. Universal Triggering Electronic Unit (UTEU) Module
Gating Prescription
Specify the prescription at TPS/OIS level
Your Treatment Planning System (TPS) and/or Oncology Information System (OIS)
allows you to specify in the prescription if gating is to be used for this treatment plan
or this treatment beam. This information is transferred to the PTS via DICOM. The
PTS will use this information to determine if gating is required or not.
Note on the OIS:
OIS EMRC integration: Gating is specified at Beam level (could vary for
different treatment beams)
OIS DEVC integration: Gating is specified at Plan level (same for all treatment
beams). In this case, the PTS extends the Gating information to each of the
individual treatment beam.
Batch/Standalone mode: It is not possible to specify gating in the prescription
when using adaPTprescribe.
Relationship between prescription and source
selected on the triggering rack
When gating is required, the PTS will assume that one of the following source must
be selected on the UTEU rack (Figure 48-2):
X-ray triggering
toggle button
Beam trigger
request LED
Power LED
Manual triggering button
Source
selector
switch