Users Manual Part 6
Clinical User’s Guide
IBA | 48-3|
Volume 1 - Treatment Session
|Part XI
- Patient Positioning and Alignment Using the Universal Beam Triggering Interface
|
For PBS, the maximal time between two requests of beam triggering (Beam
Hold request to Beam Resume request) will be greater than 3 sec and shorter
than 30 sec.
The PTS shall guarantee the following latency requirements for the PBS beam
delivery technique:
The system latency from reception of beam hold request to actual beam off
state (beam hold latency) is equal or less than 100 ms.
The system latency from reception of beam resume request to actual beam
on state (beam resume latency) is equal or less than 200 ms.
The PTS shall guarantee the following latency requirements for the DS beam
delivery technique:
The system latency from reception of beam hold request to actual beam off
state (beam hold latency) is equal or less than 120 ms.
The system latency from reception of beam resume request to actual beam
on state (beam resume latency) is equal or less than 120 ms.