Users Manual Part 6
Clinical User’s Guide
IBA | 43-15|
Volume 1 - Treatment Session
|Part IX
- adaPT
deliver
in Physics Mode Using adaPTdeliver in Physics Mode
|
Treatment
The TREATMENT SCREEN in Physics mode differs somewhat from the same screen in
Clinical mode, as follows:
Physics optional actions: this is an additional tab on the Monitor panel of the
T
REATMENT SCREEN. This tab enables you to select any given layer and to keep
track of the monitor units during your operations.
From the Irradiation Options pane you can select one of the following actions:
Pause after layer
Pause after tuning
Enable field restart
Tuning required
Display pop-ups: you can opt to display scanning controller error pop-up
messages.
The Layer Selection pane enables you to select the desired layer, as follows:
reselect the first layer
select the previous layer
select the next layer
specify any specific layer that you want to select; click the adjacent arrows
button to proceed.
Time progress bar: using this bar on the right of the screen you can keep track
of how the irradiation is evolving.
Important The time progress bar is a cumulative display. Which means that if you
irradiate the same layer a few times (e.g., when skipping, restarting,
pausing, resuming irradiation within a layer or field), the Monitor Units
(UM) do not restart from 0.
This means that in Physics mode the time progress bar is not
representative of the irradiation activities.