Users Manual Part 6

Clinical User’s Guide
IBA | 46-13|
Volume 1 - Treatment Session
|Part XI
- Patient Positioning and Alignment Introducing Patient Positioning and Alignment
|
Automatic Weight Check
Perform the automatic weight check as follows:
1. Long touch on PPS icon in hand-pendant for the POSITIONER screen. The Weight
check status is orange (Figure 46-7).
WARNING Be careful to avoid collisions of the Patient Positioning System.
Collisions may damage the PPS. It may also deteriorate the robot
accuracy and performance.
In case of a collision, the robot must be calibrated.
Important Make sure that the area near the PPS robot is clear before starting the
automatic weight check.