Users Manual Part 6
Clinical User’s Guide
IBA | 43-13|
Volume 1 - Treatment Session
|Part IX
- adaPT
deliver
in Physics Mode Using adaPTdeliver in Physics Mode
|
Be aware that the imaging system must be out of the beam path, like in Clinical
mode.
Figure 43-11. PMS Interface Panel
The differences are as follows:
To l e r a n c e s : if the values are out of tolerance, you do not need to acknowledge
any message to confirm that you are aware that the positioning devices are out
of tolerance.
Drawer: the drawer can be IN or OUT, disregarding the prescription. In Clinical
mode, the drawer must be in position to proceed to the I
RRADIATION SCREEN. In
Physics mode, this condition is irrelevant.
Note: The drawer must not be in between IN and OUT; it must be fully IN or
fully OUT).
Color legend:
Orange: screen elements in orange indicate that these settings are
acceptable in Physics mode but will block or issue an error message in
Clinical mode.
Orange, with a red frame: these settings are blocking, even in Physics
mode (see Figure 43-12, where the drawer is still in the beam path).