Users Manual Part 6

IBA | B-2 |
Clinical User’s Guide
Volume 1 - Treatment Session
Using the Dosimetry Manager
|
Starting the Dosimetry Manager
The Dosimetry Manager is accessible via a button at the bottom of the adaPTdeliver
screens. (Once clicked, the button is grayed out; see Figure B-1.)
Figure B-1. adaPTdeliver: Dosimetry Manager Button
The main display of the Dosimetry Manager is presented in Figure B-2.
The screen is divided in two parts:
one for automatic acquisition of temperature and pressure display (Automatic
TPCF mode)
one for manual input of temperature and pressure values (Manual TPCF mode).
When the automatic mode is activated, the part of the screen dedicated to the
manual mode is disabled. When the manual mode is activated, the part of the screen
dedicated to the automatic mode is disabled.
Note: For the Proteus PLUS, the Automatic TPCF mode is always disabled because
there are no sensors.
If the Automatic TPCF mode is always disabled (i.e. if the system does not include
temperature and pressure sensors), only the manual mode can be activated.