deliver
in Worklist Mode Resuming after a Partial Irradiation
|
Resuming Irradiation in the Same Session
The user can resume the irradiation in the same session when there is:
Termination by the Therapist, or
Termination by the PTS where the adaPTdeliver session is maintained
The irradiation can be resumed in the same session only after an online recovery.
Note: Resuming the irradiation in a new session is possible in both the Standalone
and the OIS mode.
1. In the BEAM SELECTION SCREEN, select the partial treatment beam manually.
Beams marked as partial are indicated by an orange checkmark .
2. In the Dose panel (Figure 41-18), the user can see that:
The delivered MU is non-zero.
The MU to be delivered is the difference between prescribed and delivered
MU.
A banner below the Dose panel gives information about the partial
irradiation.
Figure 41-18. Resuming irradiation after partial treatment