Users Manual Part 5
IBA | 41-18 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part VII
- Using
ada
PT
deliver
in Worklist Mode Resuming after a Partial Irradiation
|
Manual Recovery from DCEU values
In case the loading of the file cannot be performed, the last resort is the use of the
DCEU values.
1. Enter the delivered cumulative MUd value related to that beam, taking into
account all previous partial irradiations of that beam and click search for partial
irradiation continuation (Figure 41-16).
2. When the search is done, the signature of a physicist is required to resume the
irradiation.
Figure 41-16. Recovery after termination by PTS - Recovery from DCEU values
Resuming Irradiation
Once the irradiation data has been recovered, in case of Offline Recovery, the
irradiation can only be resumed in a new session (See section “Resuming Irradiation
in a New Session”).
Note: In case of resuming from the irradiation records containing simulated values,
a signature will be needed to proceed (Figure 41-17).