Users Manual Part 5
Clinical User’s Guide
IBA | 41-11|
Volume 1 - Treatment Session
|Part VII
- Using
ada
PT
deliver
in Worklist Mode Resuming after a Partial Irradiation
|
Figure 41-8. Recovery after termination by PTS - Manual recovery process (online)
7. In case the process is not successful, proceed to section “Manual Recovery
from DCEU values”.
Manual Recovery from DCEU values
In case the loading of the file cannot be performed, the last resort is the use of the
DCEU values.
1. Enter the delivered cumulative MUd value related to that beam, taking into
account all previous partial irradiations of that beam and click search for partial
irradiation continuation (Figure 41-9).
2. When the search is done, the signature of a physicist is required to resume the
irradiation.
a
b
c