Users Manual Part 5

IBA | 41-6 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part VII
- Using
ada
PT
deliver
in Worklist Mode Resuming after a Partial Irradiation
|
Termination by the Proton Therapy System
Online recovery - adaPTdeliver session
maintained
When the irradiation is terminated by the PTS, popup appears indicating the cause of
termination (Figure 41-3). Click OK to continue.
Figure 41-3. Recovery after termination by PTS - Warning message
The recovery of the data can be performed through one of the following methods:
1. Automatic recovery
2. Manual recovery from file
3. Manual recovery from DCEU values
Automatic recovery
1. In the popup message, type in the values of the MU delivered and dose counts
from the DCEU (Figure 41-4). Click OK to continue and proceed to section
Resuming Irradiation.