Volume 1 - Treatment Session
Chapter 41
• • • • • •
Resuming after a Partial Irradiation
Note: Screen representations throughout this chapter and throughout this part may
show ‘Control Mode: EMR-Centric’ or ‘Control Mode: DEV-Centric’ at the bottom.
When working in Standalone mode, you will notice that the Control Mode indication
reads ‘Stand-Alone’.
WARNING Whenever the system prompts the user to manually input the DCEU
value or the MUd value (MU delivered) corresponding to a partial
irradiation, the user must follow the instructions in the user
documentation and be extremely careful not to commit any typos or
input any false values. After inputting the value manually, the user must
double-check the value displayed by the system to ensure the input is
correct.
Any fault in the manually input values entails a serious risk of under or
over-irradiation of the patient.