Users Manual Part 5

Clinical User’s Guide
IBA | 40-3|
Volume 1 - Treatment Session
|Part VII
- Using
ada
PT
deliver
in Worklist Mode Operating in Worklist Mode
|
Figure 40-3. Couch Positions Check - Pop-Up Message
There is a configurable tolerance for the check (default value 2 mm), which can be
modified by IBA following a user request.
Click OK if you want to proceed with your selection, else click Cancel.
There is a user right: ISOCENTER_CHECK_OVERRIDE to authorize users to continue
with the workflow despite the consistency check finding an inconsistency.
For the configuration of the user rights, refer to Appendix BManaging PTS Users
on page B-1.
Appointment Selection: Dose Limits Check
When you select an appointment, a dose limits check is performed to verify that the
daily and total dose limit is not surpassed. You can perform as follows:
Any of the dose limits are surpassed: you can opt to override that/those limit(s)
with the appropriate user rights: there is a right for each type of limit (daily/total):
enter your user name and password and click OK. Else click Cancel.
Not all data is available to perform the dose limit checks: the OIS provides
incomplete data for the computation, the dose limits check
will be ignored. Do you want to proceed anyway? (Yes/No
message appears. Click Ye s to acknowledge or No to stop.
Grayed Out Appointments
Invalid appointments appear grayed out. Mouse (hover) over any such grayed out
appointment to see more information about the issue with that specific appointment.