Users Manual Part 5
Clinical User’s Guide
IBA | 39-23|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Resuming after a Partial Irradiation
|
Resuming irradiation
Ensure the partial beam is selected as indicated above. Click Next to proceed to the
P
REPARATION SCREEN and complete the irradiation of the selected beam.
Note: In case of resuming from the irradiation records containing simulated values,
a signature will be needed to proceed (Figure 39-20).
Figure 39-20. Confirmation to the use of simulated data for resuming irradiation
Resuming Irradiation in a New Session
The user needs to resume the irradiation in a new session when there is:
Termination by the PTS where the adaPTdeliver session crashes, or
Closing of the adaPTdeliver session by the Therapist
Note: Resuming the irradiation in a new session is possible in both the Standalone
and the OIS mode.
Using the Standalone mode
WARNING When working in Standalone mode, after a partial irradiation, the user
must record the cumulative MU value delivered for that beam
according to the center’s best practices (may be on paper or other
method). The goal is to be able to use that value for the event of a
resumption in a different session.
Not doing this entails a serious risk of under or over-irradiation of the
patient.