Users Manual Part 5
Clinical User’s Guide
IBA | 39-19|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Resuming after a Partial Irradiation
|
2. Then, click LOAD FILE (Figure 39-16a).
Note: In case the Operator cannot recover files manually, the recovery based
on the DCEU values must be performed. See section ”Manual Recovery from
DCEU values”.
3. In file explorer popup, browse and select the file as indicated by the Operator.
Click Open.
4. In the Manual Recovery popup, the button turns green and now reads File
Loaded. Click Next to continue (Figure 39-16b).
5. In case the process is successful (Figure 39-16c), click OK to continue and
proceed to section “Resuming Irradiation”.
Figure 39-16. Recovery after termination by PTS - Manual recovery process (offline)
6. In case the process is not successful, proceed to section “Manual Recovery
from DCEU values”.
a
b
c
WARNING Once the system constructs a DICOM Treatment Record and stores it on
disk as a result of the Offline Recovery procedure in a system that
functions in DEVC mode, the user needs to import the recovered
DICOM record in the OIS (with caution to avoid any double recordings).
Making a double recording in the OIS entails a serious risk of under-
irradiation of the patient.