Users Manual Part 5
Clinical User’s Guide
IBA | 39-17|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Resuming after a Partial Irradiation
|
Figure 39-13. Recovery after termination by PTS - Automatic data recovery
information (offline)
Note: In case the prescribed dose was delivered and the irradiation does not need
to be resumed, the corresponding beam can be marked off as complete using the
Mark As Complete button. The beam will be recorded in the PTS database.
3. Select the line that corresponds to the irradiation you want to recover.
Note: A value of 300 counts on the DCEU corresponds to 1 MU. See Chapter 17,
Resetting a Dose Counter Electronic Unit for details.
4. If the selected irradiation is partial, enter the corresponding DCEU values that
you have saved at the time of the partial treatment. Click OK.
WARNING The user must only proceed to mark a Beam Report as complete using
the Offline Recovery dialogue if the irradiation was completely
delivered (the prescribed dose was delivered according to the DCEU, so
the irradiation does not need to be resumed) and is correctly recorded
as complete in the OIS or the paper charts. The irradiation will be
recorded in the PTS Database using a simulation based on the specified
meterset for the irradiation and not on the actual delivered values.
This action is irreversible. Marking an incomplete Beam Report as
complete when the irradiation was not completely delivered entails a
serious risk of under-irradiation of the patient.