Users Manual Part 5

Clinical User’s Guide
IBA | 39-15|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Resuming after a Partial Irradiation
|
Offline recovery - adaPTdeliver session
crashes
When the irradiation is terminated by the PTS and the adaPTdeliver session crashes,
The Offline Recovery method is available in the SESSION DEFINITION SCREEN in
adaPTdeliver.
This screen has the button Offline Recovery (Figure 39-12), which is the starting
point of the recovery process.
WARNING When performing the offline recovery procedure, when the system
prompts the user to input the MU and counts values, the user must be
careful to input the values corresponding to the irradiation they want to
recover. The values displayed on the DCEU at that time may not be the
ones needed for the procedure (e.g., if there has been another
irradiation in the meantime). They must refer to the MU and counts
values that they noted once the irradiation stopped, following a
software issue.
Any fault in the manually input value entails a serious risk of under or
over-irradiation of the patient.
WARNING If ever a software issue occurs during an irradiation (the GUI freezes or
there is any kind of software crash), the user must take note of the
values on the DCEU (both MU and counts) once the irradiation stops.
These values will be required to recover the irradiation data offline.
Not recording these values correctly for later use entails a serious risk
of under or over-irradiation of the patient.
Important Offline recovery is mandatory as soon as adaPTdeliver software crash
occurs.
Offline recovery enables the reconciliation of available irradiation data
in the adaPTdeliver software