Users Manual Part 5
IBA | 39-12 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Resuming after a Partial Irradiation
|
Figure 39-8. Recovery after termination by PTS - Manual data recovery information
(online)
2. Call an operator to gather the irradiation data inside a .zip file in the local disk.
3. Once the operation is complete, click LOAD FILE (Figure 39-9a).
Note: In case the Operator cannot recover files manually, the recovery based
on the DCEU values must be performed. See section“Manual Recovery from
DCEU values”.
4. In file explorer popup, browse and select the file as indicated by the Operator.
Click Open.
5. In the Manual Recovery popup, the button turns green and now reads File
Loaded. Click Next to continue (Figure 39-9b).
6. In case the process is successful, click OK to continue and proceed to section
“Resuming Irradiation”.