Users Manual Part 5
IBA | 39-2 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Resuming after a Partial Irradiation
|
WARNING When working in Standalone mode, after a partial irradiation, the user
must record the cumulative MU value delivered for that beam
according to the center’s best practices (may be on paper or other
method). The goal is to be able to use that value for the event of a
resumption in a different session.
Not doing this entails a serious risk of under or over-irradiation of the
patient.
WARNING When working in Standalone mode, the user must ensure that in every
new session, beams that were partially irradiated in a previous session
are handled correctly. This means:
- Identifying a partial beam as such, taking into account the record done
of any previous partial irradiations according to the center’s best
practices (may be on paper or other method).
- Taking the appropriate actions according to the instructions in the user
documentation to continue a partial irradiation in a new session in
Standalone mode.
Not doing this entails a serious risk of under or over-irradiation of the
patient.
WARNING Whenever the system prompts the user to manually input the DCEU
value or the MUd value (MU delivered) corresponding to a partial
irradiation, the user must follow the instructions in the user
documentation and be extremely careful not to commit any typos or
input any false values. After inputting the value manually, the user must
double-check the value displayed by the system to ensure the input is
correct.
Any fault in the manually input values entails a serious risk of under or
over-irradiation of the patient.