Users Manual Part 5

IBA | 37-4 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Starting Irradiation
|
The aforementioned error message appears if the following is encountered:
Delivered MU of layer i > prescribed MU of layer i + X%
Where X is a configurable tolerance (in the above case 5% of prescribed dose); this
value may be different at your treatment center.
In case this message appears, verify that the Delivered MU on the DCEU
corresponds to both the prescribed and the recorded value. In case there is a
discrepancy, contact IBA and request to verify the calibration; otherwise, report the
warning to IBA.