As mentioned in Table 43-1 and Table 43-2, both baselined and unbaselined plans
and beams of both clinical and QA patients appear in the list.
To enable you to modify certain beam characteristics in Physics mode, the following
fields are editable:
Gantry angle
Delivered MU: any partially delivered MU appear here.
Required MU: this enables you to override the prescribed MU and rescale the
weights of the spots accordingly.
Figure 43-9. Beam Selection Screen
Preparation
The behavior on the following panels of the EQUIPMENT PREPARATION SCREEN in
Physics mode is different from Clinical mode:
PMS interface panel: refer to section “Verifying the Position of Devices” on
page 43-12.
Room interlocks: refer to section “Verifying the Room Interlocks” on
page 43-14.