Users Manual Part 5

IBA | 43-4 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part IX
- adaPT
deliver
in Physics Mode Using adaPTdeliver in Physics Mode
|
It is important to know that unlike in Clinical mode, in Physics mode error messages
and pop-up messages are prevented and treatment records are not stored.
For information generic to both Clinical mode and Physics mode, refer to Part V,
Using adaPTdeliver.
Session Definition
In Physics mode, the background color of the step indication and the step description
of adaPTdeliver screens is violet.
In addition, at login, the You logged into Physics mode. Clinical
treatments are not allowed in this mode pop-up message appears. Click
OK to continue.
Note: The PTS can be connected to the OIS. The PTS will receive the plan from the
OIS but will not take charge of the session in OIS. From then on the PTS is mastering
the workflow.
The OIS can work in QA mode or in Clinical mode.
As for adaPTinsight, the behavior of the PTS in Physics mode when connected to the
OIS will be identical to its behavior in Clinical mode, for DICOM prescriptions.
Figure 43-3. Physics Mode Acknowledgment Message