Users Manual Part 5
IBA | 36-10 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Monitoring and Controlling the Treatment
|
Monitoring Nozzle Cyclic Checks
The NOZZLE CYCLIC CHECKS TAB enables you to monitor the checks that are performed
on the status of miscellaneous nozzle components such as the DCEU, etc. You can
visually monitor the status of the check. For instance, the Beam Stop In condition is
activated only during a Pause, whereas the Beam Stop Out condition is activated only
during Irradiation.
You can access the N
OZZLE CYCLIC CHECKS TAB also by clicking the nozzle icon ( )
from the Equipment panel of the T
REATMENT SCREEN.
CAUTION In case the Beam Line Checks tab shows that one of the BAEU or BIREU
relay check is failed, request (from the maintenance team) a verification
of the structural integrity and the proper functioning of the IC cyclo
related equipment (i.e. IC cyclo related electrical wiring).
If the maintenance operation cannot be performed immediately as
recommended, it is possible to postpone the maintenance operation
and to proceed with the treatment by entering a specific password (click
on the failed check for instructions). In this case, treatment will be
delivered with degraded beam presence detection at the exit of the
cyclotron.
In any case, verification must be performed within 24 hours.