Users Manual Part 4

IBA | 35-26 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Preparing the Equipment for Irradiation
|
Figure 35-17. Room Interlocks Panel - Not Ready
When all room interlock information is available to enable irradiation, the colored
state bar reads Ready (indicated by the green color).
The R
OOM INTERLOCKS PANEL - NOT READY indicates the status of the following safety
devices:
Room master key switch: the master key is inserted in the key switch on the
MCR Control Panel and the key switch is turned on.
Room search complete: the treatment room is properly searched and the door
to the treatment room is closed.
Safety system in Treatment Mode: the TCR master key is inserted in the key
switch on the TCR Control Panel and the key switch is turned on.
Emergency stops status: all emergency stop buttons the entire PT system are
reset.
Make sure that all safety signals are present. The colored state bar reads Ready
(indicated by the green color).
CAUTION In case an error message (from the clinical Graphical User Interface
(GUI)) has warned you (while attempting to prepare a treatment beam)
that a major issue with the IC cyclo related equipment has been
detected, you shall request (from the maintenance team) a verification
of the structural integrity and the proper functioning of the IC cyclo
related equipment (i.e. IC cyclo related electrical wiring).
It will not be possible to proceed with treatment until the issue is fixed
and preparation succeeds.