Users Manual Part 4

Clinical User’s Guide
IBA | 35-21|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Preparing the Equipment for Irradiation
|
To manually enter a correction vector, click Enter correction from the PMS I
NTERFACE
PANEL and enter the corrections that are obtained from the PPVS software (i.e.,
adaPTinsight) in the corresponding fields on the B
EAM CORRECTION SCREEN that
appears.
Figure 35-12. Beam Correction Screen
Click Apply and the correction values appear on the PMS I
NTERFACE PANEL.
Validating the Positioning Parameters
Implement the positioning parameters using the wireless hand-pendant or the
remote positioning controls.
The colored state bar reads Ready (indicated by the green color). The Confirmed
status is also indicated by the green checkmark ( ) adjacent to the tabletop icon,
and the Preparation confirmed status.